Test principle
The product adopts the double antibody-sandwich method to detect antigens.
Materials and Components
(1)Test device
(2)Sample collection tube with extraction reagent
(3)Sampling kit (Optional)
(4)Instruction
(5)Timer
Product Information
specification
1pc/box; 2pcs/box; 10pcs per box
specimen
vaginal secretion
storage
2-30 ℃
Specimen requirements
1. This kit is only suitable for the detection of E7 oncoprotein expressed by HPV-16 and (or) HPV-18 in female cervical exuded cells. lt is only used for in vitro diagnosis and should not be used for the detection of other body fluids.
2. In the process of specimen collection, pay attention in strict accordance with the sampling requirements to prevent wrong detection results caused by the small number of sample cells collected.
3. When there is a large amount of visible mucus and bleeding in the specimen,the detection will be affected, resulting in wrong detection results. Therefore, it is recommended that patients avoid the period of menstruation and ovulation for sampling.
4. vaginal drugs and vaginal lubricants will affect the test, resulting in false test results. Therefore, no sexual and vaginal medication is required with in the first three days of sampling.
5. after the first collection, if you need to collect again, please collect a week later,and use new test kits.
Test Procedure
Carefully read the reagent instruction before using the test kit and strictly operate according to the instruction to ensure reliable results. Bring test device, extraction reagent and specimens were restored to room temperature 15-30℃(59-86℉).
(Note: These diagrams only show self-test operation, if professional use, please refer to the manual)
1. Testing
(1)Place the sampling device on the patient’s cervical transformation area and turn it six times clockwise.
(2)Unscrew the cap of the sample collection tube counter clockwise.
(3)Carefully place the tip of the swab of the sampling device into the sample collection tube to avoid contact between the tip of the swab and the outside of the tube.
(4)After the swab head comes into contact with the bottom of the sample collection tube, invert it10 times to allow the extraction reagent inside the tube to fully release the sample inside the extraction reagent.
(5)Place on a dry, clean and level surface and allow to stand at room temperature for 10minutes ± 5 minutes.
(6)Take out the test device from sealed foil pouch and place on a dry, clean and level surface.
(7)Unscrew the cap of the sample collection tube, hold the sample collection tube vertically. Add 2-3 full drops of the sample without air bubbles into the specimen well (S) of the test device.